AirLite 110-100空气采样泵技术参数:

SKC大气采样泵流量5~3000ml/分钟
: ± 5%
采用CPC 和低流量容器附件,可达到5~ 500 ml/分钟的低流量
体积小,重量轻
重量仅 12盎司 (340 克)
运行时间长
在大部分条件下,其运行时间超过8小时  经济且节省时间
使用性的AA 碱性电池
更换电池时不中断运行
完全整体化采样
在反压和温度改变时进行流量补偿
美国SKC采样器AIRLITE在电池电量太低时关闭以保持补偿流量
LED 显示屏指示泵运行、限定流量(流量故障)和电池状态
非光滑的RFI-处理的外壳
具备流量故障关闭选项

110-100airlite3.png

1、拧松固定带夹并拆下带夹。


2、从美国SKC采样器AIRLITE壳的背面拆下单支护螺丝。

3。轻轻地向下按压进气软管倒钩和幻灯片的堆栈和控制板的底部的情况。

4。用小艾伦扳手松开对泵壳,每边的固定螺丝

5、拧松电池盖上的螺丝,卸下底板上盖,并拆下电池。

6。分离的电池和电机线束从控制板上的电线连接器。

7。电梯控制板O?底板。

8、用钳将电机从泵壳中拔出,轻轻地摇动电机。

 Patented* flow control system holds constant flow to

± 5% of set-point

 1000 to 3000 ml/min flow range

Low flows from 5 to 500 ml/min with All-in-One low flow adapter/holder

 

 Alkaline battery powered — no charging!

 Economical, disposable AA alkaline batteries

 Ready to go anytime!

 

 Ideal for abatement, indoor air sampling, and emergency response in non-hazardous locations

 

 Minimum of 8 hours run time

Up to 3000 ml/min and 10 inches water back pressure

 

Ordering Information

AirLite requires 1/4-inch ID tubing.


Easy operation

Turn it on, calibrate, and sample

 Easy access battery compartment

 

AirLite?  fits in the palm of your hand!

Only 12 ounces (340 grams)

4.6 x 3 x 1.75 inches (11.7 x 7.6 x 4.4 cm)

 

 Features maintain sample integrity

 Constant flows up to 3000 ml/min

 Compensates up to 20 inches water back pressure at 2 L/min

 Compensates for temperature changes from 41 to 122 F (5 to 50 C)

 Low battery and flow fault shut-off and indicator

 Rugged RFI/EMI-shielded case, CE marked



1.Unscrew the two screws retaining the belt clip and remove belt clip.

2.Remove the single retaining screw from the back of the pump case.

3.Press down gently on the inlet hose barb and slide the stack and control board out of the bottom of the case.

4.Use a small Allen wrench to loosen the setscrew on each side of the pump housing

5.Loosen the screw on the battery case cover, remove cover from the base plate, and remove batteries.

6.Detach both the battery and motor wire harnesses from the wire connectors on the control board.

7.Lift the control board o? the base plate.

8.Use pliers to gently rock the motor back and forth while pulling the motor from the pump housing.

品牌/商标

美国SKC

企业类型

贸易商

新旧程度

全新

产品型号

AIRLITE