PR100手持式接收机|PR100|罗德与施瓦茨
价格:电议
地区:
电 话:86-0755-83345158
手 机:15112590728
传 真:86-0755-83356908
产品型号:
PR100

供应PR100手持式接收机|PR100|罗德与施瓦茨

R&S 罗德与施瓦茨PR100 手持式接收机

厂商名称:德国罗德与施瓦茨Rohde&Schwarz R&S

  • Ergonomic and rugged design for portableuse
  • Detection of unknown signals
  • Frequency scan and memory scan
  • Location of emissions using theR&S® HE300 portable directionalantenna
  • Reliable location of sources ofinterference
  • Demodulation and audio monitoring ofemissions
  • Signal storage in instrument andexternally
  • Signal demodulation at up to 500 kHzbandwidth
  • 15 IF filters from 150 Hz to 500kHz
  • RF spectrum with user-definable frequencyrange
  • IF spectrum with continuous display range from 10kHz to 10 MHz
  • Realtime bandwidth of 10 MHz
  • Spectrum and waterfall display
  • Output of complex baseband data
  • 6" VGA color display
  • LAN interface (SCPI) for remote control andoutput of measurement data and I/Q data

Specificationsin Brief

RF section
Frequencyrange 9 kHz to 7.5GHz
Inputimpedance 50 Ω
VSWR ≤3:1
Inputlevel typ. 0dBm
Inputattenuation typ. 10 dB(manually switchable)
Preselection
9 kHz to 30MHz 30 MHz lowpassfilter
20 MHz to 1.5GHz tuned bandpassfilters
1.5 GHz to 7.5GHz twohighpass/lowpass filter combinations
Noisefigure (preamplifierswitched on)
9 kHz to 30MHz typ. 16dB
30 MHz to 1.5GHz typ. 14dB
1.5 GHz to 3.5GHz typ. 14dB
3.5 GHz to 7.5GHz typ. 18dB
IP3(input) (preamplifierswitched off, test signal offset ≥1 MHz)
9 kHz to 30MHz typ. +20dBm
30 MHz to 1.5GHz typ. +12dBm
1.5 GHz to 3.5GHz typ. +18dBm
3.5 GHz to 7.5GHz typ. +18dBm
Phasenoise typ. 90 dBc/Hz(∆f = 10 kHz, fc = 500 MHz)
IF section
IF demodulationbandwidth 15 filters(specification of 3 dB bandwidth)
150 Hz, 300 Hz,600 Hz, 1.5 kHz, 2.4 kHz, 6 kHz, 9 kHz, 15 kHz, 30 kHz, 50 kHz, 120kHz, 150 kHz, 200 kHz, 300 kHz, 500 kHz
IFpanorama FFT, internal(2048 points), max. 20 pictures/s
Displayrange 10 kHz to10 MHz
Control
Squelch –30 dBµV to 130dBµV (in 1 dB steps)
Gaincontrol AGC,MGC
Frequencycontrol AFC
automaticretuning of signals of unstable frequency
±½ IF bandwidth(150 Hz to 500 kHz)
Signal processing
Demodulationmodes AM, FM, PULSE,I/Q (all IF bandwidths), USB, LSB, CW, ISB (only IF bandwidths ≤15kHz)
Scan
Frequencyscan
Start/stopfrequency user-selectable
Stepwidth user-selectable
Panoramascan
Start/stopfrequency user-selectable
Stepwidth 125 Hz, 250 Hz,500 Hz, 625 Hz, 1.25 kHz, 2.5 kHz, 3.125 kHz, 6.25 kHz, 12.5 kHz,25 kHz, 50 kHz, 100 kHz
Scanspeed max. 200 FFTsper second
Memoryscan
Memorylocations 1024 channels,user-programmable
Measurement accuracy anddisplays
Frequencyresolution 1Hz
Frequencyaccuracy
Versusspecified temperature range ≤1ppm
Versusinstrument aging ≤1ppm/year
Signallevel –137 dBm to 0dBm
–30 dBµV to 107dBµV
0.1 dBresolution
Displayerror max. ±3 dB /typ. ±1.5 dB
Levelindication modes AVG, FAST,PEAK, RMS
Offset up to ±½ IFbandwidth (150 Hz to 500 kHz)
1 Hzresolution
ADC 14bit
Interfaces
Antennainput
9 kHz to 7.5GHz N female /50 Ω
Max. level(nondestructive) +20 dBm / 0 VDC
Referenceinput
10MHz BNC female /50 Ω
Max. level(nondestructive) 0 dBm to +10dBm / 0 V DC
IFoutput
21.4MHz BNC female /50 Ω
uncontrolled
I/Qoutput LAN (bandwidth≤500 kHz)
Video output,digital LAN (bandwidth≤500 kHz)
Audio output,digital LAN (bandwidth≤12.5 kHz)
Audio output,analog 10...300 Hz to12.5 kHz
dependent on IFfilter and modulation
stereo jack /100 Ω
0 V to ≥2V
Data andcontrol interfaces LAN (Ethernet10/100BaseT)
USB 2.0 (toread out the SD card)
Power supply
AC (withexternal power supply) 100 V AC to 120V AC
200 V AC to 240V AC
50 Hz / 60Hz
DC 15 VDC
Battery Li-Ion, 6cells
Mechanical data
Dimensions(height x width x depth) 320mm x 192 mm x 62mm
Weight typ. 3.5kg
Environmental data
Operatingtemperature range 0°C to +50°C
Permissibletemperature range –10 °C to +50°C