特点
交直流耐压、绝缘、接地、接触电流、电气性能以及低压启动一键完成
完全符合IEC、UL、GB等安规要求
负载容量达8KVA/30Arms,满足大容量负载测试
可选购RS-232接口
可搭配华仪电源或1931系列产品使用
100%台湾研发制造,品质有保证
选购信息
编号7128多功能综合安规测试仪-产品及附件线上询价
7128ComplianceElectricalSafetyTester
1150DUT电源输出连接线
1151DUT电源输入连接线
1152大电流测试回路线(40A)
1153大电流测试线(40A)
INPUT | ||||
Voltage | 115 / 230Vac ± 15% | |||
Frequency | 50 / 60Hz ± 5% | |||
Fuses | Fuse 6.3A / 250V Slow-Blow | |||
Range | Resolution | Accuracy | ||
AC WITHSTAND VOLTAGE | ||||
Output Voltage, KVAC | 0-3.50 *1 | 0.01 | ± (2% of setting + 5V) | |
Output Frequency | 50Hz/60Hz ±0.1%, User Selection | |||
Output Waveform | Sine Wave , Crest Factor = 1.3 - 1.5 | |||
Output Regulation | ± (1% of output + 5V), From no load to full load | |||
SETTINGS | ||||
HI and LO-Limit current, mA | 0.00-40.00 | 0.01 | ± (2% of setting + 2 counts) | |
Ramp Timer, second | 0.1 - 999.9 | 0.1 | ± (0.1% of setting + 0.05 sec) | |
Dwell Timer, second | 0, 0.2 - 999.9 (0=continuous) |
|||
Current Offset, mA | 0.00-2.00 | 0.01 | ± (2% of setting + 2 counts) | |
DUT Output | ON/OFF BY User Selection | |||
Arc Detection(Option) | 1-9 ranges (9 is the most sensitivity) | |||
DC WITHSTAND VOLTAGE | ||||
Output Voltage, KVDC | 0 - 4.00 | 0.01 | ± (2% of setting + 5V) | |
DC Output Ripple | <5 % at 4KV/5mA | |||
SETTINGS | ||||
HI and LO-Limit current, mA | 0.00 - 5.00 | 0.01 | ± (2% of setting + 2 counts) | |
Ramp Timer, second | 0.1- 999.9 | 0.1 | ± (0.1% of setting + 0.05 sec) | |
Dwell Timer, second | 0, 0.2 - 999.9 (0=continuous) |
|||
Current Offset, mA | 0.00 - 0.20 | 0.01 | ± (2% of setting + 2 counts) | |
DUT Output | ON/OFF BY User Selection | |||
Arc Detection(Option) | 1-9 ranges (9 is the most sensitivity) | |||
Discharge Time | No capacitive Load is<30 msec, capacitive Load is<100 msec. | |||
Maximum Capacitive Load DC Mode | 1µF < 1KV,0.5µF < 3KV,0.75µF < 2KV, 0.08µF < 4KV | |||
INSULATION RESISTANCE | ||||
Output Voltage, DCV | 100 - 1000 | 1 | ± (2% of reading + 5 volts) | |
SETTINGS | ||||
HI-Limit, resistance, MΩ | 0,1 - 1000 (0 = OFF) | 1 | Same as Resistance MEASUREMENT Accuracy | |
Lo-Limit, resistance, MΩ | 1 - 1000 | |||
Delay Timer, second | 0, 0.5 - 999.9 (0=continuous) |
0.1 | ± (0.1% of setting + 0.05 sec) | |
GROUND BOND | ||||
Output AC Current, A | 3.0 – 30.0 | 0.1 | ± (2 % of setting + 0.02A) | |
Output Voltage, Vac | AC 6 volts Max. | |||
Output Frequency, Hz | 50Hz/60Hz ±0.1%, User Selection | |||
Milliohm Offset | 0-100 | 1 | ± (2 % of setting + 2 counts) | |
SETTINGS | ||||
HI and LO-Limit Resistance, mΩ | 0 – 510 (3.0-10.0A) | 1 | ± (2 % of setting + 2 counts ) | |
0 – 200 (10.1-25.0A) | ||||
0 – 150 (25.1-30.0A) | ||||
Dwell Timer, second | 0, 0.5 - 999.9 (0 = continuous) |
0.1 | ± (0.1% of setting + 0.05 sec) | |
MEASUREMENT | ||||
Range | Resolution | Accuracy | ||
Voltage, KV(AC/DC) | 0.00 - 4.00 | 0.01 | ± (1.5 % of reading +1 count) | |
Voltage, Vdc (IR only) | 0 - 1000 | 1 | ± (1.5% of reading + 2 counts) | |
AC Current, mA | 0.00 - 40.00 | 0.01 | ± (2% of reading + 2 counts) | |
DC Current, mA | 0.00 – 5.00 | 0.01 | ||
AC Current, A (G.B) | 0.0-30.0 | 0.1 | ± (3 % of reading + 3 counts) | |
Resistance, MΩ (IR) | 1-999.9 | 0.1/0.01 | 100-499V ± (7% of reading + 2 counts) 500-1000V ± (3% of reading + 2 counts) |
|
Resistance, mΩ (GB) | 0-510 | 1 | ± (2 % of reading + 2 counts) | |
RUN TEST | ||||
DUT POWER | ||||
AC Voltage | 0 - 277.0V, Single phase unblance, 0 - 30.0A maximum | |||
Current | 30A maximun continuous | |||
Power Rating | 8400W maximum | |||
DUT Protection | Short Circuit current 50A < 3 sec, Inrush Current 180A Response time 10u sec. | |||
SETTINGS | ||||
HI and LO Limit AC Voltage, V | 0.0-277.0 | 0.1 | ± (1.5% of setting + 0.2 V), 60.0-277V | |
HI and LO Limit AC Current, A | 0.0-30.0 | 0.1 | ± (2% of setting + 2 counts) | |
HI and LO Limit AC Power, W | 0 - 8400 | 1 | ± (5% of setting + 6 counts) | |
HI and LO Limit Power Factor | 0.000-1.000 | 0.001 | W/VA, calculated and displayed to three significant digits | |
HI and LO Limit Leakage Current | 0.0-10.00, Hi-Limit 0=OFF |
0.01 | ± (2% of setting + 2 counts) | |
Delay Time, second | 0.2-999.9 | 0.1 | ± (0.1% of setting + 0.05 sec) | |
Dwell Time, second | 0, 0.1-999.9 (0=continuous) | |||
MEASUREMENT | ||||
Range | Resolution | Accuracy | ||
Voltage, Vac | 0.0-277.0 | 0.1 | ± (1.5% of reading + 2counts), 60.0~277V | |
Current, Aac | 0.0-30.0 | 0.1 | ± (2% of reading + 2counts) | |
Power, Watts | 0-8400 | 1 | ± (5% of reading + 6 counts) | |
Power, Factor | 0.000-1.000 | 0.001 | W/VA, calculated and displayed to three significant digits | |
Leakage Current, mA | 0.00-10.00 | 0.01 | ± (2% of reading + 2 counts) | |
MD Circuit | Leakage Current measuring resistance = 2K Ω ±1% | |||
Timer, second | 0.0-999.9 | 0.1 | ± (0.1% of reading + 0.05 sec) | |
TOUCH CURRENT TEST | ||||
DUT | ||||
DUT Input Power Rating | 0 - 277.0V, Single phase unblance, 30A maximum | |||
Current | 30A maximun continuous | |||
DUT Protection | Short Circuit current 50A < 3 sec, Inrush Current 180A Response time 10u sec. | |||
SETTINGS | ||||
HI and LO Limit, µA | Range 0-6000µA, Resolution 1µA,Hi-Limit 0=OFF | |||
Delay Time, second | Range 0, 1.0-999.9 (0=continous), Resolution 0.1second. | |||
Compliance Standards | A UL544 NP | |||
B IEC60990 Fig4 U2 | ||||
X External MD(1K Ohm) | ||||
F Frequency check, External MD(1K Ohm) | ||||
Probe settings | G-L, G-N | |||
CURRENT MEASUREMENT | ||||
Frequency Range | DC-100KHz | |||
Touch current Range (RMS) | Resolution | Accuracy (DC,15-100KHz) | ||
0-6000µA | 1 | ±(2% of reading + 3counts) | ||
Touch current offset | Resolution | Accuracy (DC,15-100KHz) | ||
0-2000µA | 1 | ±(2% of reading + 3counts) | ||
LINE VOLTAGE MEASUREMENT | ||||
Range | 0.0-277.0 VAC | |||
Resolution | 0.1 V | |||
Accuracy | ± (1.5% of reading +0.2V), 60.0-277.0V | |||
External MD | Basic measuring element 1K Ω | |||
MD A - D components | Accuracy: Resistance ± 1% Capacitance ± 5% | |||
MD Voltage Limit | Maximum 30V peak or 30VDC | |||
External Terminal on TC, allowing for High, Low, Open settings (Option) | Add a extemal chanel to reach P-G, S-G and P-S for hipot test. | |||
Extemal chanel | HV relay | GND relay | ||
P-G | Open | ON | ON | |
S-G | High | OFF | ON | |
P-S | Low | ON | OFF | |
GENERAL | ||||
PLC Remote Control |
Input : Test, Reset, Interlock, Recall Memory 1-6 |
|||
Output:- Pass, Fail, Test-in-Process |
||||
Memory |
20 memories, 10 steps, step and step can link. |
|||
Remote Output | provide the +24V, GND and I/O1~7 to control 67xx and 68xx AC Power Source or Transformer Box 1931 | |||
Display | 20 x 2 LCD | |||
LED display | BUS REM(WITH RS232), LOCK | |||
Interface | Optional RS232 | |||
Security | Key Lock capability to avoid unauthorized access to all test parameters. Memory Lock capability to avoid unauthorized access to memory locations. | |||
Fail stop | This allows the 7128 to be configured to continue testing after detecting a failure.All pass/fail results are displayed at the completion of the test cycle. | |||
Alarm Volume Setting | Range : 0~9 ; 0=OFF, 1 is softest volume, 9 is loudest volume. | |||
Calibration | Software and adjustments are made through the front panel. | |||
Environment | 0-40 ℃ , 20-80%RH | |||
Dimensions/Net Weight | 430mm(W)×133mm(H)×500 mm(D)/22Kg | |||
STANDARD ACCESSORIES | ||||
Power Cord(10A) | x1 | |||
Fuses | ×2 (Including a spare contained in the fuse holder) | |||
Interlock Disable Key(1505) | x1 | |||
Hipot Test Lead, 1.5m (1101) | x1 | |||
Ground Bond Test Lead 40A, 3m(1153) | x1 | |||
Ground Bond Return Lead 40A, 3m(1152) | x1 | |||
LLT Receptacle Adaptor Box 30A , 1.6m(1932) | x1 | |||
DUT Power Input Lead 40A, 3m(1151) | x1 | |||
DUT Power Output Lead 40A, 3m(1150) | x1 | |||
*1 4.00KVAC without TUV/CE listed can be available for option |