安规综合分析仪7440
价格:电议
地区:广东省
电 话:0769-82307098
手 机:15338034669
传 真:0769-82307098
图文详情
产品属性
相关推荐
● 产品特点 | |||||||||||
☆交/直流耐壓、絕緣阻抗、接地阻抗測試 | |||||||||||
☆20×2點矩陣LCD顯示器 | |||||||||||
☆可內建4組或8組高壓矩陣式掃描通道 | |||||||||||
☆內建50組程式記憶、每一組8個測試步驟的連接測試 | |||||||||||
☆測試線路阻抗自我歸零校正功能,保證測試結果的準確性 | |||||||||||
☆鍵盤鎖定功能 | |||||||||||
☆內建PLC Remote遙控介面 | |||||||||||
☆可外接標準GPIB/RS232/Printer Card作控制 | |||||||||||
● 技术参数 | |||||||||||
MODEL | 7410 | 7420 | 7430 | 7440 | |||||||
AC WITHSTAND VOLTAGE | |||||||||||
Output Rating | 5KVAC/30mA | 5KVAC/30mA | 5KVAC/40mA | 5KVAC/40mA | |||||||
Range | Resolution | Accuracy | |||||||||
Output Voltage,ACV | 0-5000 | 0.01 | ±( 2% of setting +5V ) | ||||||||
Output Frequency | 50Hz/60Hz±100ppm,User Selection | ||||||||||
Output Waveform | Sine Wave,THD<2%( Resistion Load ),Crest Factor = 1.3-1.5 | ||||||||||
Output Regulation | ±( 1% of output +5V ) | ||||||||||
SETTINGS | |||||||||||
Hi-Limit AC Current,mA | 0-30.00 | 0-30.00 | 0-40.00 | 0-40.00 | |||||||
Resolution:0.001,Accuracy:±( 2% of setting +3counts ) | |||||||||||
Lo-Limit AC Current,mA | 0-9.999 | ||||||||||
Resolution:0.01,Accuracy:±( 2% of setting +3counts ) | |||||||||||
Ramp Time,second | 0-999.9 | 0.1 | ±( 0.1% + 0.05sec ) | ||||||||
Dwell Time,second | 0,0.3-999.9 | 0.1 | |||||||||
( 0=continuous ) | |||||||||||
Arc Detection | 1-9 ranges ( 9 is the most sensitivity ) | ||||||||||
AC Current Offset,mA | 0-2.000 or Auto Set | ||||||||||
DC WITHSTAND VOLTAGE | |||||||||||
Output Rating | 6KVDC/9999uA | ||||||||||
Output Voltage,DCV | 0-6000 | 1 | ±( 2% of setting +5V ) | ||||||||
Output Ripple | < 5% ( 6KV/9999uA at Resistive Load ) | ||||||||||
SETTINGS | |||||||||||
Hi-Limit DC Current,uA | 0-9999 | 1 | ±( 2% of setting +3counts ) | ||||||||
Lo-Limit DC Current,uA | 0-999.9 | 0.1 | ±( 2% of setting +3counts ) | ||||||||
Ramp Time,second | 0.1-999.9 | 0.1 | ±( 0.1% + 0.05sec ) | ||||||||
Dwell Time,second | 0,0.4-999.9 | 0.1 | |||||||||
( 0=continuous ) | |||||||||||
Arc Detection | 1-9 ranges ( 9 is the most sensitivity ) | ||||||||||
DC Current Offset,uA | 0-200,0 or Auto Set | ||||||||||
Ramp High Current | ON/OFF,User Selection | ||||||||||
Charge-Low Current,uA | 0-350,0 or Auto Set | ||||||||||
Discharge Time | ≦200msec | ||||||||||
MODEL | 7410 | 7420 | 7430 | 7440 | |||||||
INSULATION RESISTANCE | |||||||||||
Output Rating | 1KVDC/9999MW | ||||||||||
Output Voltage,VDC | 100-1000 | 1 | ±( 2% of setting +3V ) | ||||||||
Hi-Limit Resistance,MW | 0.1-9999(0=OFF ) | 1 | ±(2% of setting +1counts) at 500-1000VDC,1-1000MW ±(8% of setting +1counts) at 500-1000VDC,1000-9999MW ±(8% of setting +1counts) at 100-500VDC,1-1000MW | ||||||||
Lo-Limit Resistance,MW | 1-9999 | 1 | |||||||||
Delay Time,second | 0,0.5-999.9 | 0.1 | ±( 0.1% + 0.05sec ) | ||||||||
( 0=continuous ) | |||||||||||
Charge-Low Current,uA | 0-3.500 or Auto Set | ||||||||||
GROOUND BOND(7440 ONLY) | |||||||||||
Output Rating | 30A/600mW ,8V | ||||||||||
Output AC Current | 3.00-30.00 | 0.01 | ±( 2% of setting +3counts ) | ||||||||
Output AC Voltage | 3.00-8.00 | 0.01 | ±( 2% of setting +3counts ) at Open circuit |
||||||||
Output Frequency | 50Hz/60Hz±100ppm,User Selection | ||||||||||
SETTINGS | |||||||||||
Hi-Limit Resistance,mW | 0-150( 3-30A ) | 1 | ±( 2% of setting +2counts ) | ||||||||
Lo-Limit Resistance,mW | 0-600( 3-10A ) | ||||||||||
Lead Resistance,Offset mW | 0-200 | 1 | ±( 2% of setting +2counts ) | ||||||||
Dwell Time,second | 0,0.5-999.9sec ( 0=continuous ) | 0.1 | ±( 0.1% + 0.05sec ) | ||||||||
GENERAL | |||||||||||
Input Voltage AC | 115/230VAC±15%,50/60Hz±5%,Fuse 5A Slow for7410/7412 model,6.3A Slow for7420/7430/7440 model | ||||||||||
PLC Remote Control | Input: Test,Reset,( Memory1.2.3 | ||||||||||
Output: Pass,Fail,Processing | |||||||||||
※Ordering Information※----------------------------------------------- | |||||||||||
7410 Dielectric Analyzer (ACW,DCW,IR) | |||||||||||
7420 Dielectric Analyzer with Build-in 8CH Scanner(ACW,DCW,IR) | |||||||||||
7430 Dielectric Analyzer(ACW,DCW,IR) | |||||||||||
7440 Electrical Safety Compliance Analyzer(ACW,DCW,IR,GB) |