图文详情
-
产品属性
-
相关推荐
英国TTi ST124/ST265 射频微波功率计
Satori ST series USB linked Power Meters - 12.4GHz, 18.5GHz or 26.5GHz
ST124SMA, ST185SMA and ST265SMA | ||
More Information: Larger Image Data Sheet |
产品规格: |
ST124 | ST265 | |
频率 | 10MHz → 12.4GHz | 10 MHz → 26.5 GHz |
测量 | -50 → +20dBm CW | -50 → +20dBm CW |
工作温度 | 0 → 50 °C | 0 → 50 °C |
冲击 | 25G, 11 ms | 25G, 11 ms |
震动 | 15G, 100 → 2000Hz | 15G, 100 → 2000Hz |
线性错误 25 +/- 5 °C | +/-0.26 dB; -40dBm → +10dBm; 50MHz ≤ f ≤ 12.4GHz+/-0.46 dB; +10dBm → +20dBm; 50MHz ≤ f ≤ 12.4GHz +/-0.43 dB; -40dBm → +10dBm; f < 50 MHz +/-1.46 dB; +10dBm → +20dBm; f < 50 MHz | ±0.27 dB; -35dBm → +20dBm; 10 MHz → 26.5 GHz |
线性变化 25 +/- 25 °C | +/-0.5 dB; 50MHz ≤ f ≤ 12.4GHz +/-0.7 dB; 10 MHz ≤ f < 50MHzUSBHygroPalm | ±0.2 dB; 10 MHz → 6 GHz; -35dBm to +20dBm ±0.25 dB; 6 GHz → 18.5 GHz; -35dBm → +20 dBm ±0.46 dB; 18.5 GHz → 26.5 GHz; -35 dBm → +20 dBm |
校正因子错误 25 +/- 5 °C | +/-0.17 dB; 10 MHz < f < 12.4 GHz | ±0.16 dB; 10 MHz → 6 GHz ±0.19 dB; 6 GHz → 18.5 GHz ±0.21 dB; 18.5GHz → 26.5 GHz |
校正因子变化 25 +/- 5 °C | ±0.45 dB; f ≥ 50MHz ±0.67dB; f < 50MHz | ±0.2 dB; 10 MHz → 6 GHz ±0.3 dB; 6 GHz → 18.5 GHz ±0.35 dB; 18.5 GHz → 26.5 GHz |
零设置 | +/-1.8 nW | ±1.6 nW |
噪声 | 0.15 nW RMS | 0.14nW 标准偏差 |
输入 SWR - 值 | 1.2:1 | 1.25:1 10 MHz → 18 GHz 1.35:1 18 GHz → 26.5GHz |
测量速度 - 50 至 - 35 dBm - 35 至 - 20 dBm - 20 至 +20 dBm | 每秒测量 8 次 每秒测量 33 次 每秒测量 50 次 | 每秒测量 8 次 每秒测量 33 次 每秒测量 50 次 |
输入功率(损坏水平) | 200mW CW (+23dBm) | 200mW CW (+23dBm) |
连接器类型 | SMA (m) | K 型* (m) |
*K Connector 是 Anritsu 公司的注册商标。 |
英国TTi
贸易商
全新
英国