供应TDA-4B 气溶胶发生器
价格:电议
地区:北京市
电 话:010 84652037-8002
手 机:13466617820
品牌美国ATI型号TDA-4B 气溶胶发生器
类型呼吸性粉尘采样仪

TDA-4B
气溶胶发生器

 

     - 这是一款新型的Tda-4B体积还小并价格低廉的气溶胶发生器。即适用于HEPA的过滤性能试验,也适用于使用尘埃粒子计数器进行的泄漏试验,只需要提供清洁的压缩空气即可进行工作。具有三个调节阀,通过调整调节阀即可实现所发生浮游浓度的调整。  技术参数: 

可使用的流量范围

50~8,100cfm (1.4-229m2min)

发生浓度

10ug/L: 流量8,100cfm(约230m2/min时)

100ug/L:流量810cfm(约23m2/min时)

测试

100ug/L:流量810cfm(约23m2/min时)

发生粒子

PAO、DOP、多分散

发生方法

1-6Laskin Nozzles

压缩气体

3-18cfm (85-510L/min) 20psi (0.14Mpa)

电 源

不需要

外形尺寸

约280x230x250mm

重量

约7.3kg

 **适用20psi(0.14Mpa)的压力时 注:不建议采用DOP  英文详细介绍: Description

ATI manufactures portable Laskin nozzle aerosol generators which produce a sub-micron polydispersed oil mist aerosol in concentrations from 10 to 100 micrograms per liter (ug/l) at air flows from 50 to 8,100 cfm.

Aerosol generators and photometers are used to integrity test or locate leaks in high efficiency air filtration systems. Filter manufacturers use this equipment to scan ULPA and HEPA filters to verify they are free from manufacturing defects. Filter certifiers use this equipment to insure that filters were not damaged in shipping and have been installed properly, eliminating any leakage.

With the proper generator and photometer combination, filter deficiencies such as pinholes, thin spots, gasket leaks, frame leaks or seal problems can be quickly and quantifiably pinpointed and corrected thus protecting product and personnel.

Both the TDA-4B and the TDA-4BLite feature several improvements including rugged stainless steel construction, large fill port, individual nozzle control, and a 3" standard sanitary outlet fitting. An optional hose adapter, part #, is available for introducing the aerosol into positive pressure systems.

Features

The TDA-4B is the latest design in rugged, lightweight Laskin nozzle generators from ATI. The TDA-4B is a small, compact aerosol generator which requires only a supply of clean, compressed air to create polydispersed sub-micron oil aerosol.

The TDA-4B has 6 Laskin nozzles. Three valves permit the unit to be operated with 1 to 6 nozzles, providing a wide range of aerosol concentrations.

The TDA-4B is recommended for testing systems with air flows of 8,100 cfm and lower. It is ideal for work stations, Negative Pressure Filtration Units, biosafety cabinets, ceiling modules, small or portable cleanrooms, or HEPA filter units in installations where an adequate supply of clean, compressed air is readily available.

 

TDA-4B SPECIFICATIONS

  • Aerosol Output Range: 50-8,100 cfm
  • Aerosol Concentration 100 ug/l @ 810 cfm*
  • Aerosol Concentration 10 ug/l @ 8,100 cfm*
  • Generator Type: 1 to 6 Laskin Nozzles
  • Compressed Air: 3 to 18 cfm @ 20 psi
  • Aerosol Type: Polydispersed (Cold)
  • Size: 10" L ?11" W ?9" H (25cm L ?28cm W ?23cm H)
  • Weight-Pounds (lbs): 16 lbs.
  • Weight-Kilograms (kg): 7.3 kg
  • Electrical: (Not Required)


Applications

  • Clean Rooms
  • Laminar Flow Benches
  • Biosafety Cabinets
  • Glove Boxes
  • HEPA Vacuum Cleaners
  • HVAC Systems
  • HEPA Filters
  • ULPA Filters
  • Negative Pressure Filtration Units
  • Surgical Suites
  • Nuclear Filter Banks
  • Collective Protection Filters

Research and Development

Industries

  • Pharmaceutical
  • Food Processing
  • Veterinary Medicine
  • Laboratories
  • Asbestos Abatement
  • Heavy Metal Processors
  • Nuclear Power and Fuel
  • Hospitals
  • Chemical Processing
  • Certifiers
  • NBC Defense Agencies
  • Governmental Labs
  • Filter Manufacturers
  • Micro Electronic
  • Electronic
  • Biotechnology
  • Micromechanics

*If more than 20 psig is used, the output concentration will increase and conversely,

if less than 20 psig is used, the output concentration will decrease.

* If using more than 2 nozzles, upstream concentration should be measured