NS-120 溶氧电极
价格:电议
地区:上海
电 话:86 021 54159271
手 机:13818045065
传 真:86 021 54159271
NS-120 溶氧电极的详细信息
    NS-120溶 氧 电 极·典 型 应 用
        广泛应用于水质处理、水文监测、废水处理、游泳池、鱼塘和化肥、化工、生物等领域的含氧量监测·测 量 原 理
        NS-120型氧电极是一种隔膜式极谱(clark)型传感器。由传导性电解质连接的一个阴极和一个阳极组成。阴极与阳极间适当的极化电压可在阴极上将O2还原:
        阴极反应:O2 + 2H2O + 4e → 4OH-
        阳极反应:4Ag + 4Cl- → 4AgCl 4e
        这些化学反应产生与所测量溶液中O2的分压成正比的电流。·特 点
        ★反应灵敏,稳定性好
        ★无需要更换膜及电解液
        ★适用于各种恶劣水质,膜不易损坏、抗污染·技 术 参 数
        测量原理:电流测量/极谱法
        检测极限: ﹤40ppb含氧量
        响应时间:25℃时含氧量下降90%﹤2min
        残余电流:无氧介质中残留的室温信号﹤满度的2%
        温度范围:测量时0~60℃
        压力范围:0~4bar
        电极信号:25℃时20~80nA(按客户要求定制)
        温度补偿:热敏电阻阻值按客户要求定制
        电极尺寸:直径12mm,长度120mm(PG13.5螺纹接口),可按客户要求定制
        电极材质:316L不锈钢壳体,POM膜帽
        透气膜:钢筛网增强的聚四氟乙烯及聚矽氧烷膜。NS-120 dissolved oxygen electrode
    * Typical Applications
        Is widely used in water treatment, hydrological monitoring, waste water treatment, swimming pools, fish ponds and fertilizer, chemical, biological and other fields to monitor the oxygen content of* Measurement principle
        NS-120 type oxygen electrode is a diaphragm-type polarographic (clark) sensor. Connected by a conductive electrolyte, a cathode and an anode component. The cathode and anode polarization voltage between the appropriate O2 reduction at the cathode will be:
        Cathode reaction: O2 2H2O 4e → 4OH -
        Anodic reaction: 4Ag 4Cl-→ 4AgCl 4e
        These chemical reactions produce the solution with the measurement of O2 partial pressure is proportional to current.* Features
        ★ responsive, good stability
        ★ No need to replace the membrane and electrolyte
        ★ applicable to all kinds of poor water quality, membrane difficult to damage, anti-pollution* Technical parameters
        Measuring principle: current measurement / Polarography
        Detection limit: <40ppb oxygen content
        Response time: 25 ℃ when the oxygen content decreased 90% <2min
        Residual current: oxygen-free medium at room temperature residual signal "full-scale 2% of
        Temperature range: 0 ~ 60 ℃ measurement
        Pressure Range: 0 ~ 4bar
        Electrode signal: 25 ℃, when 20 ~ 80nA (according to customer request)
        Temperature Compensation: thermistor resistance customized according to customer requirements
        Electrode size: diameter 12mm, length 120mm (PG13.5 threaded interface), according to the customer's requirement
        Electrode Material: 316L stainless steel shell, POM membrane cap
        Breathable membrane: Steel mesh enhanced PTFE and polydimethylsiloxane membrane.