大龙TopPette 手动单道移液器 广州东南仪诚
价格:150.00
地区:广东省 广州市
电 话:020-61139815
手 机:15975615427

   轻便且设计合人机工效学; ? 数字视窗,令所设定量程一目了然;
 ? 量程范围广,0.1μL至10mL;
 ? 使用附件工具,能方便快捷地进行校准和维修; ? 准确的分液,每支移液器都遵照EN/ISO8655标准进行校准; ? 下半支可高温高压。

量程增量测量体积(μl)允许大系统误差
(不准确度)
允许大随机误差
(不jing确度)





%μl%μl
0.1-2.5μl0.05μl2.52.50%0.06252.00%0.05
1.253.00%0.03753.00%0.0375

0.2512.00%0.036.00%0.015

0.5-10μl0.1μl101.00%0.10.80%0.08
51.50%0.0751.50%0.075

12.50%0.0251.50%0.015

2-20μl0.5μl200.90%0.180.40%0.08
101.20%0.121.00%0.1

23.00%0.062.00%0.04

5-50μl0.5μl500.60%0.30.30%0.15
250.90%0.2250.60%0.15

52.00%0.12.00%0.1

10-100μl1μl1000.80%0.80.15%0.15
501.00%0.50.40%0.2

103.00%0.31.50%0.15

20-200μl1μl2000.60%1.20.15%0.3
1000.80%0.80.30%0.3

203.00%0.61.00%0.2

50-200μl1μl2000.60%1.20.15%0.3
1000.80%0.80.30%0.3

501.00%0.50.40%0.2

100-1000μl5μl10000.60%60.20%2
5000.70%3.50.25%1.25

1002.00%20.70%0.7

200-1000μl5μl10000.60%60.20%2
5000.70%3.50.25%1.25

2000.90%1.80.30%0.6

1000-5000μl50μl50000.50%250.15%7.5
25000.60%150.30%7.5

10000.70%70.30%3

2-10ml0.1ml10ml0.60%600.20%20
5ml1.20%600.30%15

2ml3.00%600.60%12

量程测量体积(μl)允许大系统误差
(不准确度)
允许大随机误差
(不jing确度)




%μl%μl
5μl5μl1.3%0.0651.2%0.06
10μl10μl0.8%0.080.8%0.08
20μl20μl0.6%0.120.5%0.1
25μl25μl0.5%0.1250.3%0.075
50μl50μl0.5%0.250.3%0.15
100μl100μl0.5%0.50.3%0.3
200μl200μl0.4%0.80.2%0.4
250μl250μl0.4%1.00.2%0.5
500μl500μl0.3%1.50.2%1.0
1000μl1000μl0.3%3.00.2%2.0
2000μl2000μl0.3%6.00.15%3.0
5000μl5000μl0.3%150.15%7.5