MODEL | 7314 |
GROUND BOND |
Output AC Current | 1~40A |
Output AC Voltage | 8V, Maximum |
Output Frequerncy | 50Hz/60Hz±100ppm, User Selection |
SETTINGS |
Max-Limit Range | Resistance | 0~600mΩ | 0~200mΩ | 0~150mΩ |
AC Current | 1.00~10.00A | 10.01~30.00A | 30.01~40.00A |
Resolution | 1mΩ |
Accuracy | Same as Resistance Measurment Accuracy |
Min-Limit Range | Resistance | 0~600mΩ | 0~200mΩ | 0~150mΩ |
AC Current | 1.00~10.00A | 10.01~30.00A | 30.01~40.00A |
Resolution | 1mΩ |
Accuracy | Same as Resistance Measurment Accuracy |
Lead Resistance Offset | Range | 0~100mΩ, Auto/manual |
Resolution | 1mΩ |
Accuracy | ±(2% of setting +2 counts) |
Dwell Time | Range | 0, 0.5~999.9sec (0=continuous) |
Resolution | 0.1 sec |
Accuracy | ±(0.1% + 0.05sec) |
MEASURMENT |
AC Current | Range | 1.00~40.00A |
Resolution | 0.01A |
Accuracy | ±(3%of reading +3count) |
Resistance | Range | 0~600mΩ | 0~200mΩ | 0~150mΩ |
1.00~5.99A 6.00~10.00A | 10.01~30.00A | 30.01~40.00A |
Resolution | 1mΩ |
Accuracy | ±3% of reading +3counts) 1.00~5.99A ±3% of reading +2counts) 6.00~10.00A
| ±3% of reading +2counts) |
GENERAL |
Input Voltage AC | 115/230Vac±15%, 50/60Hz ± 5%, Fuse 6.3A |
PLC Remote Control | Input: Test , Reset,Withstand Processing ,(Memory1.2.3, Interlock for 7314/7315/7316 model only) |
Output: Pass, Fail, Processing, Reset Out, Start Out |
Memory | 10 memories, 3 steps/memory |
Display | 128×64 Graphic LCD |
Key Lock | To prevent unauthorized alteration of the test Parameters |
Calibration | Build-in software and external calibrated meters |
Verification | Build-in software driven verification menu to test fault detection circuits |
Alarm Volume Setting | Range: 0-9 ;0=OFF, 1 is softest volume, 9 is loudest volume. |
Environment | 0-40℃, 20-80%RH |
Dimension | 215(W)×89(H)×430(D)mm |
Net Weight | 14kg |
STANDARD ACCESSORIES |
Power Cord(10A) | ×1 |
Fuses | ×2 (Includeing a spare contained in the fuse holder) |
Ground Bond Reutrn Lead 40A, 1.6m(1137) | ×1 |
Ground Bond Tester Lead 40A, 1.6m(1138) | ×1 |