MZ马赫-曾德尔干涉仪
价格:电议
地区:
电 话:86 0571 88076956
手 机:13758185506
传 真:86 0571 88077926

Mach-Zehnder 马赫-曾德尔干涉仪

特性

  • 两种型号可供选择:中心波长为850纳米或者1300纳米
  • 适合扫频源输出频率监测,带平衡探测输出
  • 低插入损耗
  • 平坦的波长响应
  • 集成的信号探测,用于功率监测和k-时钟信号
  • 紧凑型设计
SpecificationsINT-MZI-850INT-MZI-1300
Optical
Interferometer Wavelength Range780-925 nm1225-1375 nm
Free Spectral Range103.3 GHz ± 5%
Fiber TypeNufern 780HP™Corning SMF-28™
Input/Output ConnectionFC/APC Pigtail, 50 cm Long
Insertion Loss* 
Electrical
Detector Material/TypeSi/PINInGaAs/PIN
Detector Wavelength Range320-1000 nm800-1700 nm
Typical Responsivity (max.)0.53A/W1.0A/W
MZI Output Bandwidth (3dB)DC-200 MHz
Conversion Gain Power Monitor**30 V/W (± 25%)60V/W (± 25%)
Conversion Gain MZI Output**30 V/W (± 25%)60V/W (± 25%)
Saturation Power100 mW @ 850 nm50 mW @ 1300 nm
Maximum Input Power250 mW (Photodiode Damage Threshold)
Electrical OutputSMA
Impedance50 Ω
DC-Offset< ± 5 mV
Power Supply± 12V, 200 mA
General
Size120 x 80 x16 mm

*包括输入和输出尾纤的接头损耗,在中心波长处测量。
**使用高阻抗负载,相对于输出功率的转换增益测量结果。当阻抗为50欧姆时,该值减半。

工作原理:

迅速扫频激光光源通常用正弦调谐元件来实现OCT成像应用所需的光学频率扫描速度。需要准确和可靠的OCT信号重新校准,这样终数据点在频率上才是等间距的。当使用可调谐激光器作为输入时,Thorlabs的MZI系列干涉仪时钟箱组件设计用来产生周期性(正弦)的K时钟信号。为了得到K时钟信号,正弦信号的值和小值在频域必须等间隔(见下面的图1)。两个值之间的差值是由自由光谱响应决定的,对于MZI系列该差值是103.3GHz。

图1:MZI时钟信号

下面的图2是该系统的原理图。内部光纤网络利用5%的输入光产生功率监测和K时钟信号。剩余的光传输到输出尾纤。特殊设计的耦合器用于为传输信号和功率监测信号实现平坦的波长相应(如下面的图3和图4)。MZI的两个信号都是用超低噪声的放大平衡光电探测器探测的。

图2:MZI系列时钟箱原理图

 

图3:INT-MZI-1300从输入到输出的耦合比

 

图4:INT-MZI-1300从输入到功率监测的耦合比