- 产品品牌:
- 泰仕Tes
- 产品型号:
- Y09-3050
TDA-2H
数字式光度计(过滤器检漏仪)
|
- 本仪器体积小.重量轻.便于携带,是当今新的变携式数字光度仪,亦称过滤器检漏仪 - 该仪器由微电脑处理器进行控制,通过操作七个按键开关即可实现仪器参数及相关测试功能的设定,气溶胶DOP.PAO的系数也均可在本体内进行设定 - 本体直接显示测试数据 - 扫描探头为选择件,探头本身也具有数据显示功能,即便是在测试现场远离本体时也可通过探头本身方便清楚的读取测试数据 - 如购买该仪器的专用测试软件,即可直接将测试数据传送至计算机中。
|
技术参数:
显示范围
|
0.0001%-100%(6位自动转换)
|
测试范围
|
0.0001%-125ug/L
|
测试
|
满量程的0.05%
|
显示(本体、扫描探头)
|
数据显示泄露率从0.0001%-100% 并同时直观显示数值变化图表
|
特殊功能
|
自动归零功能、报警功能
|
数据输出
|
RS—232C
|
流量
|
1cfm
|
电源
|
100~250VAC,50/60Hz
|
外形尺寸
|
约257×147×363mm
|
重量
|
本体
|
约7kg
|
扫描探头(可选件)
|
1.1kg
|
英文详细介绍:
Description
The 2H is the latest digital photometer available in a portable self-contained unit. The 2H has been designed with the user in mind for simplicity of operation and virtually maintenance-free use. Its basic purpose is to detect and measure mass concentration of aerosol. Three-stage airflow sensor and optics indicator lights improve data reliability.
The microprocessor-controlled 2H enables the operator to quickly set operating functions and parameters. It is easily programmed and operates with just 7 push buttons. The digital display has a set-up menu and diagnostic codes for operator convenience. The range of concentration measurable is from 100 to 0.0001 µg/l. The 2H has two pre-set reference values (DOP & PAO) for aerosol agents as well as a USER setting, which allows storage of a sampled concentration to be retained as a reference value.
2H Portable Photometer Specifications
SIZE: 10.1” x 14.3” x 5.8”
25.7 cm x 36.3cm x 14.7 cm
WEIGHT: 15.5 lbs. plus 2.5 lbs. for scanning probe and accessories = 18.0 lbs 7.0 kg plus 1.1 kg for scanning probe and accessories = 8.2 kg.
ENCLOSURE: Instrument enclosure is rugged, die cast aluminum and can be easily disassembled as required for recalibration and service of internal components. The enclosure has a swiveling handle that also acts as a support, permitting tilting of the control panel display for easy viewing, as well as increasing the ease of transport.
POWER: 100 to 250 volts AC, 50/60 Hz fused power entry module with spare fuses. Automatically adjusts to virtually all standard voltages used worldwide. Please specify power supply cable when ordering unit.
AUTO-RANGING: Automatic from 0.0001% to 100.0%. Provides 6 decades of display and eliminates the manual range selection switch.
AUTO ZERO: Automatically establishes zero by pressing the 0% button when sampling the airflow through the internal ULPA reference filter in the clear mode. Automatically establishes downstream reading as zero by pressing the 0% button when sampling the downstream airflow.
ALARM: Audible alarm in unit sounds when adjustable set point is exceeded. Visible alarm flashes on control panel and scanning probe. Audible portion of alarm can be disabled.
BARGRAPH: Solid-state bar graph with high visibility green LED on control panel and scanning probe.
AMPLIFIER: Solid-state linear amplifier.
PHOTOMETER ACCURACY: 1% of full scale for the decade in use in ratio metric mode.
DIGITAL DISPLAY: Shows % Leakage from 0.0001% to 100.0%
DYNAMIC RANGE: 0.0001 to 125 micrograms per liter. Improved firmware prevents system errors by blanking the display when sampling above 125 micrograms per liter; display is restored when concentration is lowered. "HIGH" mode enables sampling of concentrations exceeding 200 micrograms per liter.
LIGHT SOURCE: Long lasting, stable, solid-state scattering chamber light source. Photometer operates at low power, generating low heat, and low stray light values. Life expectancy is 50,000 to 100,000 hours.
REPEATABILITY: 0.5% full scale
SCATTERING CHAMBER: Field-tested design for efficient and reliable performance. A three-stage “OPTICS” indicator ensures accurate internal reference response; green indicates internal reference within 5% of original value, orange indicates internal reference shifted to 5-10% of original value, and red indicates that the internal reference has shifted by more than 10%.
VACUUM PUMP: True 1 cubic foot per minute (cfm) sample rate (28.3 liters per minute) using a twin head, carbon vane pump for quiet operation and dependability. A HEPA filter is externally integrated at the outlet of the vacuum pump entrapping any particles from the pump exhaust. A “FLOW” indicator ensures operation within required parameters.
OUTPUT: A DB9 connector is provided standard on all units for RS232 output