分析盒技术数据 |
探头种类
|
温度测量 |
O2 测量
|
CO (H2 补偿) |
低CO 测量 |
CO2
|
NO 测量 (350M可选择)
|
低NO 测量
|
NO2measuring module (350M可选择) |
SO2 测量 |
量程 |
-40... +1200°C |
0... +25 Vol % O2 |
0... +10000 ppm CO |
0... +500 ppm CO |
0... CO2 max Vol. % CO2 |
0... +3000 ppm NO |
0... +300 ppm NO |
0... +500 ppm NO2 |
0... +5000 ppm SO2 |
|
± 0.5% of mv (+100... +1200°C) ±0.5°C (-40... +99.9°C) |
±0.8% of f.v (0... +25 Vol. % O2)
|
± 5% of mv (+100... +2000 ppm CO) ± 10% of mv (+2000... +10000 ppm CO) ± 5 ppm CO (0... +99 ppm CO) |
± 5% of mv (+40... +500 ppm CO) ± 2 ppm CO (0... +39.9 ppm CO) |
由O2计算 |
± 5% of mv (+100... +1999.9 ppm NO) ± 10% of mv(+2000... +3000 ppm NO) ± 5ppm NO (0... +99 ppm NO) |
± 5% of mv (+40... +300 ppm NO) ± 2 ppm NO (0... +39.9 ppm NO) |
± 5% of mv (+100... +500 ppm NO2) ± 5 ppm NO2 (0... +99.9 ppm NO2) |
± 5% of mv (+100... +2000 ppm SO2) ±10% of mv (+2001... +5000 ppm SO2) ±5 ppm SO2 (0... +99 ppm SO2) |
分辨率 |
0.1C (-40... +1200°C) |
0.1 Vol% O2 (0... +25 Vol.% O2) |
1 ppm CO (0... +10000 ppm CO) |
0.1 ppm CO (0... +500 ppm CO) |
0.01 Vol % CO2 |
1 ppm NO (0... +3000 ppm NO)
|
0.1 ppm NO (0... +300 ppm NO) |
0.1 ppm NO2 (0... +500 ppm NO2) |
1 ppm SO2 (0... +5000 ppm SO2) |
反应时间 |
|
20 s |
40 s |
40 s |
20 s |
30 s |
30 s |
40 s |
30 s |
反应类型 |
|
t 95 |
t 90 |
t 90 |
t 95 |
t 90 |
t 90 |
t 90 |
t 90 |
探头种类
|
效率 |
Flue gas loss
|
压差 1 |
压差 2 |
流速 |
|
|
|
|
量程
|
0... +120%
|
-20... +99.9% qA
|
-200... +200 hPa
|
-40... +40 hPa
|
0... +40 m/s
|
|
|
|
|
|
|
|
± 1% of mv (-50... -200 hPa) ± 1 % of mv (+50... +200 hPa) ± 5hPa (-49.9... +49.9 hPa) |
± 1% of mv (-40... -3 hPa) ± 1% of mv (+3... +40 hPa) ±0.03 hPa (-2.99... +2.99 hPa)
|
|
|
|
|
|
分辨率 |
0.1% (0... +120%)
|
0.1 % qA (-20... +99.9qA)
|
0.1 hPa (-200... +200 hPa)
|
0.01 hPa (-40... +40 hPa)
|
|
|
|
|
|
尺寸 |
395 x 275 x 95 毫米 |
重量
|
3200 克 |
存放温度
|
-20... +50°C |
操作温度 |
-5... +45°C |
外壳物料 |
ABS |
其它技术参数 |
记忆: 250000 读数 电力供给: via intergrated mains unit (90 V to 260 V. 47 to 63 Hz) or exchangeable rechargeable batteries 供电要求: 0.5 A (110 AC), 0.3A (230 V AC) 露点计算: 0 至99°C td 正压: 50 hPa (500 毫米水注) 负压: 200 hPa (2000 毫米水注) 泵流量: 0.8m/s (设有流量监测) 尘埃: 20 g/m³ dust in flue gas 湿度: +70°C 入气口露点 Measuring range extension (dulition) for CO: dilution factors 0, 2, 5, 10, 20 40 烯析气体: 新鲜空气或 N2 度: Reading plus max. 2% 保养: 分析盒: 2年(不包括工作零件及传感器) O2 传感器:1 ½ 年 探头: 1年(不包括过滤器) 充电池: 1年 配件: ½ 年 打印机: 1年
|