Current Probes SS-270/SS-260 SS-270 SS-260
Specifications and performance Model | SS-270 | SS-260 |
---|
Frequency Bandwidth | DC to 2MHz(-3dB) | DC to 10MHz(-3dB) |
---|
Maximum input current | Maximum input range | 500A rms | 150A rms |
---|
Maximumpeak current | 700A peak, non-continuous | 300A peak, non-continuous |
---|
Output sensitivity | 0.01V/A |
---|
Sensitivity accuracy | ±1.0% of reading ±5mV for probes only (in the range from 0 to 500A rms, DC, and AC 45 to 66Hz) | ±1.0% of reading ±1mV for probes only (in the range from 0 to 150A rms, DC, and AC 45 to 66Hz) |
---|
±2.0% of reading for probes only (700A peak or less and over 500A rms, DC, and in the range of AC 45 to 66Hz) | ±2.0% of reading for probes only (300A peak or less and over 150A rms, DC, and in the range of AC 45 to 66Hz) | Noise | 25mA rms or less (observed with an 20MHz oscilloscope) |
---|
Power consumption | 7.2VA max | 5.5VA max |
---|
Measureble wire diameter | ?20mm |
---|
PS-26 Power Supply for SS-250/SS-240A/SS-270/SS-260 | This is a current probe power supply for use when performing measurements with an oscilloscope that does not have an active probe power supply. | Specifications | Conforming current probes | SS-250/SS-240A/SS-240 | Number of power connectors | 2 | Output voltage | ±12 V ±0.5 V | Output current | 600 mA max. (total of all channels) | Operating temperature and humidity | 0°C to +40°C, 80% RH or less (no condensation) | Storage temperature and humidity | -10°C to +50°C, 80% RH or less (no condensation) | Operating elevation | Up to 2000 m, indoors | Power supply voltage | AC 100 V (Can be set to 120, 220, or 240 V) (taking into account a voltage variation of ±10% for the power supply voltage) | Rated power supply frequency | 50Hz / 60Hz | Power consumption | 20 VA max. | Size | Approx. 73 (W) x 110 (H) x 186 (D) mm | Weight | Approx. 1.1 kg (except accessories) | Complied with standards | Safety | EN61010-1:2001 Pollution degree: 2 | EMC | EN61326:1997+A1:1998+A2:2001 EN61000-3-2:2000 EN61000-3-3:1995+A1:2001 | Accessories | Power cord (1), instruction manual (1), spare fuse (1) ?When shipped with 100/120 V product: F1.0 AL/250 V (f5 x 20 mm) for 100/120 V ?When shipped with 220/240 V product: F0.5 AL/250 V (f5 x 20 mm) for 220/240 V |
| * | With the power supply for current probe PS-26, if the current being measured is large, it may not be possible to use two SS-250/SS-240A units at the same time. |
|