肌松监测仪 美国
价格:电议
地区:
电 话:86 010 62913283
手 机:13001184827
传 真:86 010 62913283

肌松监测仪(爱尔兰)

爱尔兰 TOF-Watch SX肌松监测仪具有以下特点:

1、有自动探测传感器和刺激电极,大而清晰的显示界面,低电流模式可适用于持续或间断的肌松监测,提供了所有的刺激模式:单次刺激(ST)、四个成串刺激(TOF)、强直刺激(TET)、强直刺激后计数(PTC)、双短强直刺激(DBS).稳定电流设计保证刺激的可靠性,即便皮肤的电阻发生改变.校正功能也能简化刺激模式中对照颤搐高度的设定.

2、标配表面温度探头和神经定位功能;

3、随机术后专用分析软件。

TOF-Watch SX肌松监测仪配置单

编号 名称 数量

01 肌松监测仪主机 1

02 刺激电极 1

03 加速度传感器 1

04 神经刺激探头 1

05 拇指固定器 1

06 数据传输缆线和接 1

07 软件安装盘 1

08 操作手册 1

09 温度探头 1



英文资料:

TOF-Watch SX is the most advanced device in the TOF-Watch range of neuromuscular transmission monitors. The instrument incorporates all features required for routine daily use. Fully compliant with Good Clinical Practice guidelines, it is equally suited for use in clinical studies.

NEW Optional Feature!
Now compatible with Phillips medical systems monitors. Click to learn more.

Flexible use
TOF-Watch SX can be used as a peripheral nerve stimulator, as a nerve locator for local anesthesia or as a neuromuscular monitor. And TOF-Watch SX can also be linked to a PC for real-time display of data that can later be exported and printed for record keeping.

Features
TOF-Watch SX is programmed to accommodate every type of stimulation pattern currently in use:

Single twitch (ST) - at 0.1 or 1 Hz
Train Of Four (TOF)
Slow TOF (user programmable time intervals from 1 to 60 min.)
Tetanic stimulation (at 50 or 100 Hz)
Post Tetanic Count (PTC)
Double Burst stimulation (DBS 3.3 and 3.2)
These patterns can also be adjusted manually using the "P button."

A constant current feature guarantees reliable stimulation even with increased skin resistance (up to 5 kOhm).

TOF-Watch SX has two automatic calibration modes, one at a user predetermined current and one which establishes the supramaximal value of the stimulation current. Manual gain adjustment is also available.

A superficial temperature sensor completes the package.

TOF-Watch SX presents all relevant data (twitches, body temperature, transducer sensitivity, stimulation mode, etc.) on an easy-to-read LCD display.

TOF Link
TOF Link, a fiber-optic cable facility, enables on-line data "dumping" from TOF-Watch SX to a PC without reciprocal electromagnetic interference with other OR or ICU appliances. During operations, data can be logged using the TOF-Watch SX Monitor program. In this way, detailed and authentic operation records can be created.

The TOF-Watch SX Monitor software makes a real-time data accessible for further processing on a personal computer. Its two basic functions are:

Logging and display of real-time recorded data on personal computers
Exporting and printing of previous recorded data
Good Clinical Practice conformity
During on-line data dumping, standard pre/programmed comments or other user comments can be added to patient recordings. Such comments can later be completed or changed. Program use is password controlled: the password can only be assigned by a main user.



Microprocessor-controlled Custom designed
LCD with numerical presentation of twitch data


Automatic Calibration:
Automatic adjustment for optimal gain and 100% reference

Stimulation Modes:
TOF (Train of Four)
PTC (Post Tetanic Count)
1 Hz twitch
0.1 Hz twitch
DBS 3.3 and 3.2 (Double burst)
Tetanic stimulation (Burst), 5 sec - 50 Hz or 100Hz

Output:
(accuracy ± 5% of display value)
Surface electrodes
Constant current, 0 - 60 mA (0 - 12 µC up to 5 kOhm
Monophasic, 200 µs pulse width
Needle electrodes
Constant current, 0 - 6 mA or (0 - 0.24 µC)
Monophasic, 40 µs pulse width

Transducer:
Acceleration transducer (accuracy ±5% of full scale value)

Battery:
9V Alkaline or NiCd (6LR61 / 6AM6)

Supply voltage:
9V=, 100 mA max

Type:
BF

Weight:
Approximately .55 lbs. with battery

Dimensions:
8 x 3.6 x 2 inches

Standard
Stimulation cable for surface electrode
Acceleration transducer
Thermistor probe for TOF-Watch SX only

Optional
Stimulation cable for local anesthesia
Mounting bracket
Hand adapter
Arm board
TOF-Link including software for TOF-Watch SX only