General Specifications (All Modes) Output Signal Maximum Output: ? 10V . Oscilloscope Input: Suitable for an input impedance of 1MOhms in parallel with < 30pF. Trace Position: Wide range DC offset control within control box. Safety Max. Circuit Voltage: 300V Cat II (on AC line circuits). 600V Cat 1 (on uncategorised secondary circuits inside equipment). Max. Tip Temperature: 150oC maximum allowable temperature at probe tip. Conformance: Complies with EN61010-1 & EN61010-031 Bandwidth Control Switch Position: Full 500kHz 2Hz Nominal Bandwidth: DC to 5MHz DC to 500kHz DC to 2Hz Risetime: <70ns 700ns 175ms Aberrations: <+/-5% <+/-1% <+/-1% Noise (Typical)*: 6mA rms 3mA rms 1.5mA rms * This is the noise level for current measurement using the Toroid attachment. For PCB track measurement the equivalent noise will depend upon the track width and gain setting but will be similar to the Toroid measurement figure for a track width of 0.5mm. HF Performance (bandwidth switch at Full) Propagation Delay: 60ns typical (to 10%). Bandwidth: DC to 5MHz (small signal) Slew Rate: 15A/us (equivalent). Overload Indication Indicator Threshold: Indicator LED within control box will light if output voltage exceeds +/- 10V or if large magnetic fields cause the system to saturate. Power Source Power Supply: 5.2V at up to 5 watts from AC line adaptor (supplied). Mechanical Probe Dimensions: 155mm x 38mm x 28mm max; 2.8mm x 1.8mm at tip Cable Length: 2m from probe tip to output BNC EMC Conformance: Complies with EN61326 Magnetic Field Measurement (Mode = Field) Scaling Factor: 250uT (or 200A/m) per Volt. Accuracy and Linearity: +/-3% Maximum Field: +/-2.5mT (2000A/m). Current Measurement using Toroid (Mode = Wire) Scaling Factor: 1 Amp per Volt. Accuracy and Linearity: +/-5% Current Range: +/-10mA to +/-10A (DC + peak) Max. Wire Diameter: 3.5mm (unbroken) or 6mm (end fed) Current Measurement in PCB Tracks (Mode = PCB Track) Scaling Factor: Adjustable to 1 Amp per Volt for track widths 0.2mm to 3.5mm (0.007" to 0.14") and 2 Amp per volt for track widths 3mm to 6.5mm (0.125" to 0.25") using Calibrator and compensation graph. Sensor Spacing: 0.7mm distance from sensor to PCB track set by probe design. Calibrator: Built-in calibrator within the control box providing an AC or DC calibration current through a 0.5mm track. |