InfraCal Soot Meter | Model HATR-CP |
Analytical Wavelength | 3.9μm, 2553cm-1 |
Sample Surface | Cubic Zirconia ATR Crystal Scratch resistant, impervious to acids |
Display | 4 digit, 7-segment red LED, 5/8 in. character height |
Power Requirements | 12V DC, 2.0A, 12V battery pack or adapter available |
Power Supply | Universal AC/DC (supplied as standard) |
Weight | 2.0kgs |
Dimensions | 165×165×225mm3 |
Suggested Operating Range | 4°C-45°C |
Typical Measurement Range | 0-15% Soot |
Output Resolution | 0.0001 |
Analysis Time | Less than 30 seconds |
Communication Port | RS232 port, 9-pin D-sub Female, PC compatible |
Instrument Repeatability | 0-3% soot ±0.1%, 3-15% soot ±0.3% |
Non-volatile Memory Calibration | Factory: 0-10% soot |
User: Up to 20 point curve fitting calibration, plus zero |
Part Number | Descriptions |
405-0011 | InfraCal Soot Meter Plus, Model HATR-CP. Complete with HATR Sample Stage for direct measurement of suspended carbon (soot) level in diesel engine oils (factory calibrated to 10% soot concentration), Power Supply and Instruction Manual. |
Accessories | |
403-0025 | Set of 6 Reference Oils in dropper bottles, appropriate for calibration and verification of reproducibility 0-12% soot range of Model HATR-CP and Model HATR-C. This set of reference oils has been correlated with the results obtained via thermal gravimetric analysis (TGA) of fresh used diesel oil samples. |
403-0018 | InfraLog Serial Printer Prints calibration and measurement data, complete with sequence number, date and time. Includes all required cables. |
403-0013 | Carrying Case with pre-diced pluck foam, 470×370×203mm3, 3.6kgs. |
403-0003 | External 12V Battery Pack with 120V/60Hz charger (for use in North America). Includes carrying case with shoulder strap and cigarette lighter adapter cable for connection to analyzer. Provides 16 hours of continuous operation. |
403-0002 | Cigarette Lighter Adapter Cable |
403-0005 | Dust Cover |
403-0001 | Universal Switching Power Supply and Line Cord for use with InfraCal Analyzers. 90-264VAC, 50-60Hz input, regulated 12V/25W output |
403-0014 | Replacement Source/Cable Assembly |
100-1005 | InfraCal Soot Meter User's Manual |
100-1011 | InfraCal Soot Meter Plus User's Manual |
Comparative Soot Measurements On Used Diesel Oil Samples | |||||||||
FT-IR, TGA, and InfraCal Soot Meter | |||||||||
Provided by Mobil Oil Company, Paulsboro Technical Center | |||||||||
Field Test Samples | Lab Engine Test Samples | ||||||||
Soot, % mass | Soot, % mass | ||||||||
Sample ID | FT-IR | TGA | Soot Meter 1 | Soot Meter 2 | Engine/Hours | TGA | Soot Meter 1 | Soot Meter 2 | |
A1 | 1.46 | 1.4 | 1.4 | Mack T-8 | |||||
B1 | 1.59 | 1.6 | 1.6 | Hours: | 0 | 0.13 | 0.1 | 0.0 | |
B2 | 1.82 | 1.8 | 1.9 | 50 | 1.76 | 1.7 | 1.7 | ||
B3 | 1.94 | 1.9 | 2.0 | 100 | 3.13 | 3.2 | 3.1 | ||
B4 | 2.56 | 2.5 | 2.5 | 150 | 4.48 | 4.5 | 4.4 | ||
A2 | 2.74 | 2.65 | 2.6 | 2.6 | 175 | 5.16 | 5.0 | 5.0 | |
B5 | 2.68 | 2.7 | 2.7 | 200 | 5.80 | 5.8 | 5.8 | ||
B6 | 2.81 | 2.76 | 2.8 | 2.9 | 225 | 6.38 | 6.1 | 6.0 | |
B7 | 2.82 | 2.90 | 2.6 | 2.7 | 250 | 6.92 | 6.1 | 6.2 | |
B8 | 3.15 | 3.0 | 3.1 | DDC Series 60 | |||||
B9 | 3.21 | 3.2 | 3.2 | Hours: | 0 | 0.09 | 0.0 | 0.0 | |
B10 | 3.53 | 3.3 | 3.1 | 24 | 0.83 | 0.8 | 0.8 | ||
A3 | 3.54 | 3.5 | 3.6 | 48 | 1.40 | 1.4 | 1.4 | ||
A4 | 4.05 | 4.0 | 4.0 | 72 | 1.98 | 1.9 | 1.9 | ||
A5 | 4.40 | 4.3 | 4.4 | 96 | 2.52 | 2.5 | 2.5 | ||
A6 | 4.99 | 4.5 | 4.6 | 120 | 3.05 | 3.0 | 2.9 | ||
A7 | 6.12 | 5.1 | 5.1 | 144 | 3.58 | 3.5 | 3.5 | ||
A8 | 6.47 | 5.7 | 5.7 | 168 | 4.10 | 4.0 | 4.1 | ||
A9 | 7.00 | 6.0 | 6.1 | 192 | 4.57 | 4.5 | 4.6 | ||
216 | 5.03 | 4.8 | 4.9 | ||||||
240 | 5.33 | 5.0 | 5.0 | ||||||
Note: A & B series samples are different oil formulations. | Note: Soot Meter data was obtained using 2 production models. Each was calibrated to 5% soot, which accounts for measurement discrepancy in samples containing a higher percentage. |