MODEL | 7130 | 7132 | 7140 | 7142 | ||||||
AC WITHSTAND VOLTAGE | ||||||||||
Output Rating | 5KVAC/20mA | |||||||||
Range | Resolution | Accuracy | ||||||||
Output Voltage,KVAC | 0-5.00 | 0.01 | ±( 2% of setting +5V ) | |||||||
Output Frequency | 50Hz/60Hz±100ppm,User Selection | |||||||||
Output Waveform | Sine Wave,THD<2%( Resistion Load ),Crest Factor = 1.3-1.5 | |||||||||
Output Regulation | ±( 1% of output +5V ),Form no load full load | |||||||||
SETTINGS | ||||||||||
Max-Limit AC Current,mA | 0-20.00 | 0.01 | ±( 2% of setting +2counts ) | |||||||
Min-Limit AC Current,mA | 0-9.999 | 0.001 | ±( 2% of setting +2counts ) | |||||||
Ramp Time,second | 0-999.9 | 0.1 | ±( 0.1% + 0.05sec ) | |||||||
Dwell Time,second | 0,0.3-999.9 | 0.1 | ||||||||
( 0=continuous ) | ||||||||||
Arc Detection | 0,1-9 ranges ( 0=OFF,9 is the most sensitivity ) | |||||||||
DC WITHSTAND VOLTAGE(7140,7142 ONLY) | ||||||||||
Output Rating | 6KVDC/7500uA | |||||||||
Output Voltage,KVDC | 0-6.00 | 0.01 | ±( 2% of setting +5V ) | |||||||
Output Ripple | < 5% ( 6KV/7500uA at Resistive Load ) | |||||||||
SETTINGS | ||||||||||
Max-Limit DC Current,uA | 0-7500 | 1 | ±( 2% of setting +2counts ) | |||||||
Min-Limit DC Current,uA | 0-999.9 | 0.1 | ±( 2% of setting +2counts ) | |||||||
Ramp Time,second | 0.1-999.9 | 0.1 | ±( 0.1% + 0.05sec ) | |||||||
Dwell Time,second | 0,0.4-999.9 | 0.1 | ||||||||
( 0=continuous ) | ||||||||||
Arc Detection | 0,1-9 ranges ( 0=OFF,9 is the most sensitivity ) | |||||||||
Discharge Time | ≦200msec | |||||||||
7130/7132/7140/7142数位合成型耐压/绝缘测试仪 | ||||||||||
MODEL | 7130 | 7132 | 7140 | 7142 | ||||||
INSULATION RESISTANCE ( 7132,7142 ONLY ) | ||||||||||
Output Rating | 1KVDC/9999MW | |||||||||
Output Voltage,VDC | 30-1000 | 10 | ±( 2% of setting +3V ) | |||||||
SETTINGS | ||||||||||
Max-Limit Resistance,MW | 0,1-9999(0=OFF ) | 1 | ±(2% of setting +2counts) at 500-1000VDC,1-1000M | |||||||
Min-Limit Resistance,MW | 1-9999 | 1 | ±(5% of setting +2counts) at 500-1000VDC,1000-9999M | |||||||
±(8% of setting +2counts) at 30-500VDC,1-1000M | ||||||||||
Ramp Time,second | 0.1-999.9 | 0.1 | ± ( 0.1% + 0.05sec ) | |||||||
Dwell Time,second | 0,1.0-999.9 | 0.1 | ||||||||
Delay Time,second | 0,0.5-999.9 | 0.1 | ||||||||
( 0=continuous ) | ||||||||||
GENERAL | ||||||||||
Input Voltage AC | 115/230VAC±15%,50/60Hz±5%,Fuse 3.15A Fast | |||||||||
PLC Remote Control | Input: Test,Reset,Memory1.2.3,Interlock | |||||||||
Output: Pass,Fail,Processing | ||||||||||
Interface( Option ) | RS485 Interface | |||||||||
Memory | 10memories,3 steps/Memory | |||||||||
Display | 128×64 Graphic LCD | |||||||||
Safety | Built-in Smart GFI trip current 450uA max,HV shut dowm speed : <1mS | |||||||||
Key Lock | To prevent unauthorized alteration of test parameters | |||||||||
Calibration | Build-in software and external calibrated meters | |||||||||
Verification | Build-in software driven verification menu to test fault detection circuits | |||||||||
Alarm Volume Setting | Range : 0-9;0=OFF,1 is softest volume,9 is loudest volume | |||||||||
Environment | 0-40℃,20-80%RH | |||||||||
Dimension/Net Weight | 215mm(W)×89mm(H)×370mm(D)/10Kg | |||||||||
※Ordering Information※------------------------------------------------------------------------------------ | ||||||||||
7130 AC Withstand Voltage Tester | ||||||||||
7132 AC Withstand Voltage / Insulation Resistance Tester | ||||||||||
7140 AC/DC Withstand Voltage Tester | ||||||||||
7142 AC/DC Withstand Voltage / Insulation Resistance Tester |