分析盒技术数据 |
探头种类
| 温度测量 | O2 测量
| CO (H2 补偿) | 低CO 测量 | CO2
| NO 测量 (350M可选择)
| 低NO 测量
| NO2measuring module (350M可选择) | SO2 测量 |
量程 | -40... +1200°C | 0... +25 Vol % O2 | 0... +10000 ppm CO | 0... +500 ppm CO | 0... CO2 max Vol. % CO2 | 0... +3000 ppm NO | 0... +300 ppm NO | 0... +500 ppm NO2 | 0... +5000 ppm SO2 |
| ± 0.5% of mv (+100... +1200°C) ±0.5°C (-40... +99.9°C) | ±0.8% of f.v (0... +25 Vol. % O2)
| ± 5% of mv (+100... +2000 ppm CO) ± 10% of mv (+2000... +10000 ppm CO) ± 5 ppm CO (0... +99 ppm CO) | ± 5% of mv (+40... +500 ppm CO) ± 2 ppm CO (0... +39.9 ppm CO) | 由O2计算 | ± 5% of mv (+100... +1999.9 ppm NO) ± 10% of mv(+2000... +3000 ppm NO) ± 5ppm NO (0... +99 ppm NO) | ± 5% of mv (+40... +300 ppm NO) ± 2 ppm NO (0... +39.9 ppm NO) | ± 5% of mv (+100... +500 ppm NO2) ± 5 ppm NO2 (0... +99.9 ppm NO2) | ± 5% of mv (+100... +2000 ppm SO2) ±10% of mv (+2001... +5000 ppm SO2) ±5 ppm SO2 (0... +99 ppm SO2) |
分辨率 | 0.1C (-40... +1200°C) | 0.1 Vol% O2 (0... +25 Vol.% O2) | 1 ppm CO (0... +10000 ppm CO) | 0.1 ppm CO (0... +500 ppm CO) | 0.01 Vol % CO2 | 1 ppm NO (0... +3000 ppm NO)
| 0.1 ppm NO (0... +300 ppm NO) | 0.1 ppm NO2 (0... +500 ppm NO2) | 1 ppm SO2 (0... +5000 ppm SO2) |
反应时间 | | 20 s | 40 s | 40 s | 20 s | 30 s | 30 s | 40 s | 30 s |
反应类型 | | t 95 | t 90 | t 90 | t 95 | t 90 | t 90 | t 90 | t 90 |
探头种类
| 效率 | Flue gas loss
| 压差 1 | 压差 2 | 流速 | | | | |
量程
| 0... +120%
| -20... +99.9% qA
| -200... +200 hPa
| -40... +40 hPa
| 0... +40 m/s
| | | | |
| | | ± 1% of mv (-50... -200 hPa) ± 1 % of mv (+50... +200 hPa) ± 5hPa (-49.9... +49.9 hPa) | ± 1% of mv (-40... -3 hPa) ± 1% of mv (+3... +40 hPa) ±0.03 hPa (-2.99... +2.99 hPa)
| | | | | |
分辨率 | 0.1% (0... +120%)
| 0.1 % qA (-20... +99.9qA)
| 0.1 hPa (-200... +200 hPa)
| 0.01 hPa (-40... +40 hPa)
| | | | | |
尺寸 | 395 x 275 x 95毫米 |
重量
| 3200克 |
存放温度
| -20... +50°C |
操作温度 | -5... +45°C |
外壳物料 | ABS |
其它技术参数 |
记忆: 250000 读数 电力供给: via intergrated mains unit (90 V to 260 V. 47 to 63 Hz) or exchangeable rechargeable batteries 供电要求: 0.5 A (110 AC), 0.3A (230 V AC) 露点计算: 0 至99°C td 正压: 50 hPa (500 毫米水注) 负压: 200 hPa (2000 毫米水注) 泵流量: 0.8m/s (设有流量监测) 尘埃: 20 g/m³ dust in flue gas 湿度: +70°C 入气口露点 Measuring range extension (dulition) for CO: dilution factors 0, 2, 5, 10, 20 40 烯析气体: 新鲜空气或 N2 度: Reading plus max. 2% 保养: 分析盒: 2年(不包括工作零件及传感器) O2 传感器:1 ½ 年 探头: 1年(不包括过滤器) 充电池: 1年 配件: ½ 年 打印机: 1年 |