/温度校验仪表
价格:电议
地区:上海
电 话:86 021 60899686
手 机:13852896228
传 真:86 021 61638388-6633
品牌RKC型号FB100
测量范围-200-+1800(℃)
Measuring accuracy
ThermocoupleType : K, J, T, E, PLII, U, L
Less than -100°C (-148°F) : ±1.0°C (±1.8°F)
-100 to +500°C (-148 to 932°F) : ±0.5°C (±0.9°F)
More than 500°C (932°F): ±(0.1% of reading + 1 digit)
Type : N, S, R, W5Re/W26Re
Less than 0°C (32°F) : ±2.0°C (±3.6°F)
0 to 1000°C (32 to 1832°F) : ±1.0°C (±1.8°F)
More than 1000°C (1832°F): ±(0.1% of reading + 1 digit)
Type : B
Less than 400°C (752°F) : ±70.0°C (±126°F)
400 to 1000°C (752 to 1832°F) : ±(1.4°C + 1digit) [±(2.5°F + 1digit)]
More than 1000°C (1832°F): ±(0.1% of reading + 1 digit)
Cold junction temperature compensation error
±1.0°C (1.8°F) [Between 5 and 40°C (41 and 104°F)]
±1.5°C (2.7°F) [Between -10 and 5°C (16 and 41°F), and 40 and 50°C (104 and 122°F)]
RTDLess than 200°C (392°F): ±0.2°C (±0.4°F)
More than 200°F(392°F) : ±(0.1% of reading + 1 digit)
DC voltage and DC current
±(0.1% of span)

Selectable sampling time among 50ms, 100ms, and 250ms.
Selectable sampling time makes the FB Series suitable for any application ranging from pressure control requiring fast response to precise control requiring highest resolution. The selections are 100ms (factory setting) usually suitable for most of standard applications, 50ms for fast response applications, and 250ms for precise control requiring high-resolution. The PID parameters can be set in 1/10 unit which supports fast and accurate control. Ramp-to-setpoint and output-change-limiter functions are each settable for UP and DOWN so that a special heater can be controlled precisely.

250ms sampling controller50ms sampling controller



Selectable PID control algorithm
PID control algorithm is selectable in the FB Series to achieve the most precise control for various applications.
PV Derivative PID : suitable for fixed setpoint control (Factory setting)
Deviation Derivative PID : suitable for ramp control using ramp-to-setpoint function and cascade control.

PV derivative type PID controlDeviation derivative type PID control
Most adaptive to conventional fixed value controlMost adaptive to conventional fixed value control



Advanced Heat/Cool PID algorithm with Undershoot Suppression
Advanced Heat/Cool PID algorithm achieves stable control by the Undershoot Suppression (USS) function, and independent P-I-D settings for both Heat and Cool which are effective for applications with strong cooling gain. Control Response type selection is settable to avoid overshoot at start-up and against external disturbance. The FB Series also offers selection of water cool and air cool PID control algorithm.

Control response parameterUSS (Under Shoot Suppression) gain

Go to FB100 top page
Easy maintenance Plug-in
Panel space saving 74mm depth
Numerous inputs and outputs
Inter-controller Communication
Auto-temperature-rize with Learning Function
Temperature Ratio Setting
Cascade Control
Group RUN/STOP Function
Recipe Function (Multi-memory Area)
Ramp/Sork Program Control
Startup tuning
Easy parameter setup via USB loader port